101
Troubleshooting
• Connect a USB stick to the receiver through the USB
device cable provided and press the Log button again.
• Restore the default settings (by pressing the
Log+Scroll+Power buttons simultaneously) in order to 
make the internal memory the active memory. Press 
the Log button again.
If neither of these two actions resolves your problem, go to 
step 3.
Step 3. Is the
Currently Used
Memory Full?
Data logging will stop automatically or won’t start if the 
storage medium used (internal memory or USB stick) is full. 
On the General Status screen, read the remaining percentage 
of free memory (second line, last number in the line). 
1. If “0%” is displayed, then the memory used is full. Do one
of the following:
•
Change the storage medium
•
Using your field terminal, empty the memory or delete 
the files you don’t need anymore.
If neither of these two actions resolves your problem, you 
may have a malfunctioning receiver. Contact your local 
dealer or email Technical Support for assistance.
2. If the memory is not full (>0%), you may have a
malfunctioning receiver. Contact your local dealer or email 
Technical Support for assistance.
Radio Data Link Fails to Provide Base Corrections to Rover
The Data Link icon is displayed on the rover’s General Status 
screen when base corrections are received and a float or fixed 
solution is available. Next to it is the age of corrections, a 
value which should not normally exceed a few seconds when 
the data link operates smoothly.
After examining the General Status screen, you determine 
that the rover is not receiving data. Follow the outline below 
to troubleshoot this problem.
RTK Base
RTK Rover
PP Base
PP Rover
Relevant to
•