4 manual operation, 2 moving the lift – Metrohm 788 IC Filtration Sample Processor User Manual
Page 96
4 Operation
788 IC Filtration Sample Processor
88
4.4 Manual
operation
4.4.1
Turning the sample rack / Positioning the samples
Using the <Í> and <Î> keys the sample rack can be
turned one position to the left (in the counterclockwise direction)
or right (clockwise).
8
MOVE
manual operation
MOVE 1 : sample
sample,spec.1}8,
1}999
Position vessel /
Turn rack
With the MOVE command, a particular sample tube or other
vessel can be placed under the needle. With
numerical rack position as well as the predefined current sam-
ple (SAMPLE command) or the special beakers 1-8 can be cho-
sen.
For security reasons turning the sample rack is only possible
when the lift is in or above the shift position.
4.4.2
Moving the lift
The keys <Ï> and <Ð> allow upward and downward move-
ment of the lift. The lowest possible lift position is defined by the
configuration parameter
'max. lift way'
.
HOME
END
The
the upper limit.
tion 4.3.2).
9
LIFT
manual operation
LIFT: 1 : rest mm
work,rinse,shift,
special,rest,
0}125 mm
Position the lift
With the LIFT command, the lift can be run to a given position.
In addition to selecting an exact position in mm (0 - 125 mm),
the
position, rinse position, shift position, special position, rest
position = 0 mm).