6 appendix, 1 error messages – Metrohm 788 IC Filtration Sample Processor User Manual
Page 141
6.1 Error messages
788 IC Filtration Sample Processor
133
6 Appendix
6.1
Error messages
When an error occurs the execution of the active command is inter-
rupted and an error message is displayed (display blinks). This must be
acknowledged with the
If an error occurs during processing of a sample series, the 788 IC Fil-
tration Sample Processor will then be switched into the "HOLD" state.
After the cause of the error has been rectified, the sample series can be
continued with the next command in the active sequence by pressing
the
ning can also be halted with
The list of possible error messages and their causes:
Error message
Cause
* battery low
The battery for the permanent storage
of user methods must be replaced.
* wrong rack
The rack positioned is not the one
that was assigned to the method un-
der
'parameters'
.
* rack data missing
No sample rack is in position or no
rack data can be found for the sam-
ple rack that is in place.
* raise lift first
Turning of a rack could not be carried
out because the lift was below the
defined shift position.
* user memory full
The memory for the user-defined
methods is full. Before saving a new
method, methods that are not used
or used only rarely must be deleted.