3 Operating tutorial
788 IC Filtration Sample Processor
48
>rack definitions 2
work position 71 mm
x In any case don’t forget to confirm the value with
>rack definitions 2
rinse position 125 mm
x The next menu item
'rinse position'
defines the height at which
the lift must be when the needle is rinsed. As for the work position,
the value here can also be entered manually or automatically ac-
cepted. For the latter, the configuration menu must be exited by
pressing
>rack definitions 2
rinse position 105 mm
>rack definitions 2
shift position 0 mm
x The menu item
'shift position'
defines the height of the needle
when the sample rack is rotated. For the 788 IC Filtration Sample
Processor this height is set to 0 mm and cannot be changed.
x Press
>rack definitions 2
special position 0 mm
x The
'special position'
defines a further height of the lift. For the
entry proceed as for the work position.
>rack definitions 2
special position 55 mm
x The final entry in the rack configuration is the definition of the posi-
tion of the special beakers.
>rack definitions 2
>>special positions
x In the submenu
'>>special positions'
enter the positions at which
you have placed conditioning or rinsing beakers (for the 6.2041.430
sample rack the two positions
'special beaker
1 128'
and
'spe-
cial beaker 2 129'
are already set).
STOP
x The configuration can now be exited with
times and must not be re-defined every time.