4.6 Standard methods
788 IC Filtration Sample Processor
99
4.6.1 Method "PC" and "PC Seg"
Application
Processing a queue with the PC program with the PC as "Master".
Interconnection
See section 2.6.1 (without suppression) and section 2.6.3 (with sup-
pression)
Program of 788 IC Filtration Sample Processor
788 IC Filtr. Proc. 5.788.0010
parameters
method PC
number of samples: rack
>start sequence
1 CTL:Rm: INIT
>sample sequence
1 SCN:Rm : Wait1
2 SCN:Rm : Pump1 ?
3 MOVE 1 : sample
( PUMP 1.1 : 5 s)
4 LIFT: 1 : work mm
5 CTL:Rm: FILL A 1
6 PUMP 1.1 : 240 s
7 CTL:Rm: ZERO 1
8 CTL:Rm: INJECT A 1
>final sequence
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:
------------
Report header with serial number and program ver-
sion
Method name
Number of samples to be processed (entire sample
rack)
Initialize remote interface
Waiting until 732 IC Det. sends signal on rem. line 3
Waiting until 709 IC Pump runs
Move needle to sample
Only "PC Seg": aspirate air during 5 s
Place lift with needle to working position
Switch injection valve A at 733 to "Fill"
Fill sample loop with sample during 240 s
Trigger autozero at 732 IC Detector
Switch injection valve A at 733 to "Inject"
—— Settings for changer functions ——
—— Reaction to manual stop ———
Settings in the «IC Metrodata» program
Definition of IC method (analysis time, start parameters for IC instru-
ments 732, 733 and 709, etc.), additional association of this method
with the following time program for the 732 IC Detector:
Method / Equipment / Setup / Program / Text:
0.0 Puls_788_set Set remote line 3 at 732/1 to 1 (s. below)
0.1 Puls_788_reset Set remote line 3 at 732/1 to 0 (s. below)
0.2 Flag
end
Program end
Method / Equipment / Setup / Program / Remote configuration:
Puls_788_set =
**1***** Set remote line 3 at 732/1 to 1
Puls_788_reset =
**0***** Set remote line 3 at 732/1 to 0
Method / Equipment:
Start with method
Procedure at program start
1. Start 788 IC Filtration Sample Processor with
2. Create new queue in the program «IC Metrodata» and start it.