Metrohm 788 IC Filtration Sample Processor User Manual

Page 41

background image

2.4 Installation for full control with «IC Net»

788 IC Filtration Sample Processor

33

5. Change this program for 'start-dialysis.smt' as follows and save it

again:

001 Ctrl INIT

732

002 Move

sample

003 Lift work
004

Ctrl

PUMP 754 on

005 Wait 120

s

006

Ctrl

FILL B 1 / STEP

007 Wait 540

s

008

Ctrl

STEP MSM 753

009 Ctrl ZERO

1

010

Wait 60 s

Initialize remote lines at 732 detector
Move needle to sample position
Place lift with needle to working position
Start 754 Pump Unit
Rinse dialysis cell with sample solution
Stop of acceptor solution
Dialysis time
Switch 753 suppressor module to next position
Trigger autozero at 732 IC Detector
Additional dialysis time (total 600 s)

End-Dialysis.smt

6. Open system file 'dialysis.smt', again. Select Control/Connect to

Workplace and save it under the new name 'end-dialysis.smt'.

7. Change the program for 'end-dialysis.smt' as follows and save it

again:

001

Ctrl

FILL A 1

002

Ctrl

INJECT B 1

003 Wait 30

s

004

Ctrl

INJECT A 1

005

Ctrl

PUMP 754 off

Switch injection valve A at 733 to "Fill"
Transfer acceptor solution to sample loop
Fill sample loop with acceptor solution for 30 s
Switch injection valve A at 733 to "Inject"
Stop 754 Pump Unit

8. The sample queue then contains the following entries, for example:

System Ident

Vial

Chrom. No. Dialysis No.

1

start-dialysis.smt Probe

0 1

- 1

2

dialysis.smt Probe

1

1

1 2

3

dialysis.smt Probe

2

2

2 3

4

dialysis.smt Probe

3

3

3 4

5

dialysis.smt Probe

4

4

4 5

6

dialysis.smt Probe

5

5

5 6

7

end-dialysis.smt Probe

6 6

6 -

This example describes the processing of a queue with six samples.

The columns 'Chrom. No.' and 'Dialysis No.' are just given for

information. Please note that in the 788 program of 'dialysis.smt' not the

given sample position 'Vial' is used; it is increased by 1 to start dialysis

of the next sample. Therefore, all samples have to be arranged on the

sample rack in a closed line.