background image

4 Operation

788 IC Filtration Sample Processor

100

4.6.2 Method "SP" and "SP Seg"

Application

Processing a queue with the PC program «IC Metrodata» with the 788

IC Filtration Sample Processor as "Master".

Interconnection

see section. 2.6.1 (without suppression) and section 2.6.2 (with sup-

pression)

Program of 788 IC Filtration Sample Processor

788 IC Filtr. Proc. 5.788.0010
parameters
method SP
number of samples: rack
>start sequence
1 CTL:Rm: INIT
2 CTL:Rm: PUMP 752 ON
>sample sequence
1 SCN:Rm : Pump1 ?
2 MOVE 1 : sample
( PUMP 1.1 : 5 s)
3 LIFT: 1 : work mm
4 CTL:Rm: FILL A 1
5 PUMP 1.1 : 240 s
6 CTL:Rm: ZERO 1
7 CTL:Rm: INJECT A 1
8 WAIT 1200 s
>final sequence
1 CTL:Rm: PUMP R/S 1
2 CTL:Rm: PUMP 752 OFF
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:
------------

Report header with serial number and program ver-

sion

Method name
Number of samples to be processed (entire sample

rack)

Initialize remote interface
Switch on 752 Pump Unit for suppressor


Waiting until 709 IC Pump runs
Move needle to sample
Only "SP Seg": aspirate air during 5 s
Place lift with needle to working position
Switch injection valve A at 733 to "Fill"
Fill sample loop with sample during 240 s
Trigger autozero at 732 IC Detector
Switch injection valve A at 733 to "Inject"
Waiting time for recording and evaluation of the

chromatogram (must be adapted)

Switch off 709 IC Pump
Switch off 752 Pump Unit
—— Settings for changer functions ——




—— Reaction to manual stop ———


Settings for 732 IC Detector

Setting the measurement parameters.

Settings in the «IC Metrodata» program

Definition of an IC method without method associated remote control.

The waiting time

WAIT

entered under

sample sequence

in the 788 IC Fil-

tration Sample Processor must be equal to the analysis time

Duration

.

Procedure at program start

1. Create new queue in the program «IC Metrodata» and start it.
2. Enter parameter

number of samples

for 788 IC Filtration Sample

Processor .

3. Start 788 IC Filtration Sample Processor with .