2.4 Installation for full control with «IC Net»
788 IC Filtration Sample Processor
31
Settings in the «IC Net 2.1» program
There are two ways to operate an automated Metrohm IC system with
dialysis:
A) For each sample, every analysis procedure contains first a dialysis
step followed by the chromatographic separation.
The total analysis time per sample results from summation of
the single times of dialysis and separation.
B) Chromatographic separation of one sample and dialysis of the next
sample is done simultaneously.
The total analysis time per sample is reduced to the time of
chromatographic separation.
Both techniques are described on the following pages.
A) Dialysis and separation of one sample in a single procedure
1. Definition of new program parameters for the 788 IC Filtration
Sample Processor under Configuration/Control:
PUMP 754 on = ---****1******
PUMP 754 off = ---****0******
2. Program for the 788 IC Filtration Sample Processor:
001 Move
sample
002 Lift work
003
Ctrl
PUMP 754 on
004 Wait 120
s
005
Ctrl
FILL B 1 / STEP
006 Wait 480
s
007 Ctrl ZERO
1
008
Ctrl
STEP MSM 753
009
Wait 120 s
010
Ctrl
FILL A 1
011
Ctrl
INJECT B 1
012 Wait 30
s
013
Ctrl
INJECT A 1
014
Ctrl
PUMP 754 off
Move needle to sample position
Place lift with needle to working position
Start 754 Pump Unit
Rinse dialysis cell with sample solution
Stop of acceptor solution
Dialysis time
Trigger autozero at 732 IC Detector
Switch 753 suppressor module to next position
Additional dialysis time (total 600 s)
Switch injection valve A at 733 to "Fill"
Transfer acceptor solution to sample loop
Fill sample loop with acceptor solution for 30 s
Switch injection valve A at 733 to "Inject"
Stop 754 Pump Unit