7 rs232 interface, 8 external bus interface – Metrohm 788 IC Filtration Sample Processor User Manual
Page 154
6 Appendix
788 IC Filtration Sample Processor
146
6.4.7 RS232
interface
This diagnosis step tests the functionality of the serial interface.
x Prepare instrument for diagnosis (see section 6.4.2).
x If necessary, press <Ð> several times until
diagnosis
>RS232 test
x
>RS232 test
RS232 test connector ?
x Insert the 3.496.8480 test plug into the RS232 interface without turn-
ing off the instrument.
x
The test runs automatically. If no error occurs, the following appears:
>RS232 test
RS232 test ok
x Remove test plug and press
diagnosis
>external bus test
6.4.8
External Bus interface
This diagnosis step tests the functionality of parts of the internal EBUS
interface.
x Prepare instrument for diagnosis (see section 6.4.2).
x If necessary, press <Ð> several times until
diagnosis
>external bus test
x
If no error occurs, the following identification appears on the display:
>external bus test
address 0x86 type 3
x
>diagnosis
>beeper test
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)