5 mic 2, Modular anion system with chemical suppression – Metrohm 788 IC Filtration Sample Processor User Manual
Page 35
2.4 Installation for full control with «IC Net»
788 IC Filtration Sample Processor
27
2.4.5 MIC
2
The 788 IC Filtration Sample Processor is connected to an anion IC
system with chemical suppression consisting of 732 IC Detector, 733 IC
Separation Center, 709 IC Pump, 752 IC Pump Unit, and 762 IC
Interface as shown in Fig. 14 using the 6.2141.110 remote connection
cable and the 6.2134.080 RS232 cable.
762
733
709
732
PC
6.2134.100
6.2125.090
788
6.2134.080
6.2141.110
752
6.2134.090
6.2128.130
6.2115.070
6.2128.180
Fig. 14: Interconnection with anion IC system with chemical
suppression (MIC 2)
Settings in the «IC Net 2.1» program
Creation of a time program for the 788 IC Filtration Sample Processor:
001 Ctrl
INIT
732
002 Move sample
003 Lift
work
004 Ctrl
ZERO
1
005 Ctrl
FILL A 1
006 Pump 240 s
007 Ctrl
INJECT A 1
Initialize remote interface at 732
Move needle to sample
Place lift with needle to working position
Trigger autozero at 732
Switch injection valve A at 733 to "Fill"
Fill sample loop with sample during 240 s
Switch injection valve A at 733 to "Inject"
Modular anion system with chemical suppression