4.3 Methods

788 IC Filtration Sample Processor

77

parameters
>start sequence

Start sequence of a sample series

The processing sequence entered here is

executed once at

the start of a sample series. This can be useful for switching

on external devices, for example.
In this submenu up to 99 command lines can be entered as a

processing sequence (see section 4.3.3).

parameters
>sample sequence

Processing sequence for each sample

This processing sequence is

executed during the process-

ing of every sample of a series.
In this submenu up to 99 command lines can be entered as a

processing sequence (see section 4.3.3).

parameters
>final sequence

Final sequence of a sample series

This processing sequence is

executed once at the end of a

sample series. This could be the useful for switching off ex-

ternal devices, for example.
In this submenu up to 99 command lines can be entered as a

processing sequence (see section 4.3.3).

parameters
>changer settings

Settings for 788 IC Filtration Sample Processor

>changer settings
rack number 2

0}16

Rack that is assigned to the method

0

= no particular rack

This setting can force the use of a certain rack with the

method chosen. If this is not desired, the rack number

'0'

must be chosen.

>changer settings
lift rate 1 12 mm/s

3}12 mm/s

Stroke speed of lift

>changer settings
shift rate 20

3}20

Rack shift rate

Turning speed of the rack in angular degrees/second