6 Appendix

788 IC Filtration Sample Processor

136

Interfaces

Remote interface

specifications, see section 5.1

RS232 interface

specifications, see section 5.2

Safety specifications

Construction / Testing

According to IEC 1010 / EN 61010 / UL 3101-1,

protection class 1

Safety directions

The Instructions for Use include information and

warnings which must be heeded by the user to

assure safe operation of the instrument.

Electromagnetic compatibility (EMC)

Emitted interference

Standards met:

EN55011 (class B), EN55022 (class B),

EN50081-1 01.92

Immunity to interference

Standards met:

IEC801-2/IEC1000-4-2 (class 4), IEC801-3/

IEC1000-4-3 (class 2), IEC801-4/IEC1000-4-4

(class 3), IEC801-5/IEC1000-4-5 (class 2/3),

IEC801-6/IEC1000-4-6 (class 2), EN55011

(class B), EN55022 (class B),

EN50081-1/2 10.92, EN50082-1 03.97,

EN61000-3, EN61326-1 03.97

Ambient temperature

Nominal operating range

+5}+40°C (at 20}80 % atmospheric humidity)

Storage, transport

–20}+60°C (at atmospheric humidity< 50 %)

–20}+50°C (at atmospheric humidity< 85 %)

–20}+40°C (at atmospheric humidity< 95 %)

Housing

Sample changer

Metal case, multiple enameling

Keyboard

Crastin (PBTB), aluminum-steamed on the inside

Dimensions

Width

280 mm

Height

660 mm

Depth

480 mm

Weight

12.4 kg (without accessories)