5.2 RS232 interface
788 IC Filtration Sample Processor
127
5.2.7
Data transmission protocol
The RS232 interface of the 788 IC Filtration Sample Processor is con-
figured as DTE (Data Terminal Equipment) with the following technical
specifications:
x Standard
Data interface in accordance with EIA standard
RS 232C (DIN 66020, page 1), transmission pa-
rameters can be set under
'configura-
tion/RS232 settings'
, see section 4.2.1.
x Control characters C
R
DEC 13
HEX 0D
L
F
DEC 10
HEX 0A
XON
DEC 17
HEX 11
XOFF
DEC 19
HEX 13
x Max. line length
80 characters + C
R
L
F
x Mode
Full duplex (simultaneous transmission and re-
ceiving)
Restriction:
If data are first received by the interface, trans-
mission is not started until the receipt is at an
end.
x Cable
For interconnection of the 788 IC Filtration
Sample Processor with third-party devices, only
a shielded data cable may be used (e.g.
METROHM D.104.0201). The cable shield must
be perfectly earthed at both devices (pay atten-
tion to current loops: always use star earthing).
Use only connectors with sufficient shielding
(e.g. Metrohm K.210.0001 with K.210.9004).
x Cable length
max. ca. 15 m
Start
7 or 8 data bits
parity bit