Metrohm 788 IC Filtration Sample Processor User Manual
Page 60
3 Operating tutorial
788 IC Filtration Sample Processor
52
>sample sequence
8 WAIT 1200 s
x In this line a waiting time is defined that is used for the acquisition of
the chromatogram.
x The LEARN mode can also be used with the WAIT command.
<Ð>
>sample sequence
9 NOP
x An empty line with a
'NOP'
-entry (no operation) always forms the
end of a sequence.
x Exit the sample sequence with
quence.
START
x After all sample tubes have been processed, the final sequence is
executed.
>final sequence
1 CTL:Rm: PUMP R/S 1
x In this line the pump drive at the 709 IC Pump is switched off. Press
START
>final sequence
2 CTL:Rm: PUMP 752 OFF
x In this line the pump drive at the 752 Pump Unit is switched off.
Press
QUIT
>final sequence
3 NOP
x Now you have reached the end of the final sequence and have
completed the entire run of a sample series.
x By pressing
returns to the normal state.
PARAM
7
SAMPLE
START
x Now prepare some sample tubes and place them on the sample
rack. Prepare the IC system with 732 IC Detector, 733 IC Separation
Center, 709 IC Pump and 752 Pump Unit for the recording of chro-
matograms.
x Enter the number of samples to be processed () and
define the position of the first sample (SAMPLE = 1).
x Now you can start your first sample series with