Metrohm 788 IC Filtration Sample Processor User Manual
Page 78
4 Operation
788 IC Filtration Sample Processor
70
>auxiliaries
beeper: ON
ON,OFF
Acoustic warning signal
ON
Acoustic warning signal in case of error
messages
OFF
No warning signal
>auxiliaries
device label
8 ASCII characters
Device
label
Freely selectable ASCII character string for characteriz-
ing the instrument (text entry, see section 4.1.6).
>auxiliaries
program 5.788.0010
read only
Number of the program version
Display only (no entry possibility).
Please specify this number in inquiries to Metrohm.
>auxiliaries
max. lift way 125 mm
0}125 mm
Maximum stroke path for needle
The following entry becomes effective after a RESET or
switching the IC Sample Processor off and on again.
This setting for max. lift way is
important for safe op-
eration. If the value for this entry is correct, damage to
the needle can be avoided because this prevents the
needle from being driven lower than the position indi-
cated. If you use the 6.2041.430 sample rack with
6.2743.050 sample tubes, the default value of 125 mm
should be retained.
Pressing
sition of the active tower.
configuration
>rack definitions
Definition of sample racks
>rack definitions
rack number 2
1}16
Number of the rack
The number of the rack in position is automatically dis-
played here, when its configuration is already stored
and when a RESET has been performed. If the configu-
ration of another rack must be changed, its rack num-
ber must be entered and confirmed with
The rack number will be shown in the first menu line for
the subsequent entries. For more information concern-
ing sample racks, see section 4.5. The number
2
is as-
signed to the 6.2041.430 sample rack delivered.
>rack definitions 2
code 010001
6 bits
Identification code of the rack
The rack code must be unique and can only occur once
in the instrument (see section 4.5).