Metrohm 788 IC Filtration Sample Processor User Manual
Page 111
4.6 Standard methods
788 IC Filtration Sample Processor
103
Settings in the «IC Metrodata» program
1. Definition of anion method for channel 1 (analysis time, start pa-
rameters for IC instruments 732, 733 and 709, etc.), additional asso-
ciation of this method with the following time program for the 732/1
IC Detector:
Method / Equipment / Setup / Program / Text:
0.0 Puls_788_set Set remote line 3 at 732/1 to 1 (s. below)
0.1 Puls_788_reset Set remote line 3 at 732/1 to 0 (s. below)
0.2 Flag
end
Program end
Method / Equipment / Setup / Program / Remote configuration:
Puls_788_set =
**1***** Set remote line 3 at 732/1 to 1
Puls_788_reset =
**0*****
Set remote line 3 at 732/1 to 0
Method / Equipment:
Start with method
2. Definition of cation method for channel 2 (analysis time, start pa-
rameters for IC instruments 732, 733 and 709, etc.). The analysis
time
Duration
must be shorter than the analysis time for the anion
method.
Procedure at program start
1. Create queue with cation method for channel 2 in the program «IC
Metrodata» and start it.
2. Start 788 IC Filtration Sample Processor with
3. Create queue with anion method for channel 1 in the program «IC
Metrodata» and start it.