2 Installation
788 IC Filtration Sample Processor
34
2.5
Installation with compact IC systems
2.5.1
Connection to 761 Compact IC or 790 Personal IC
Compact IC systems 761 Compact IC and 790 Personal IC are remote
controlled by PC software. The 788 IC Filtration Sample Processor is
connected to these IC systems via its remote connection:
788
761 / 790
Fig. 17: Interconnection with compact IC systems
Tubing connections are described in section 2.10.2 of the instructions
for use for the 761 Compact IC and 790 Personal IC, respectively.
The 6.2141.110 remote cable is needed for connecting the 788 IC
Filtration Sample Processor to the compact IC system (761 or 790) as
follows:
Electrical connection 788 – 761/790
x Connect the end of the 6.2141.110 Cable marked with "766" to the
remote interface 30
30
30
30 788 IC Filtration Sample Processor.
x Connect the end of the 6.2141.110 Cable marked with "732/1" to
the remote interface of the 761 Compact IC and 790 Personal IC
respectively.
x The third end of this cable is not required with this configuration.
Settings at the 788 IC Filtration Sample Processor
x For operation with the 761 Compact IC or 790 Personal IC it is
recommended that the program "761" is used with the 788 IC
Filtration Sample Processor (see section 4.6.3):
6.2141.110 cable