4 installing the splash protection, 5 installing the needle or tubing – Metrohm 788 IC Filtration Sample Processor User Manual
Page 21
2.3 Attaching the accessories
788 IC Filtration Sample Processor
13
2.3.4
Installing the splash protection
To avoid any danger of injury by the needle, the 6.2751.040 splash
protection must always be installed when operating the 788 IC
Filtration Sample Processor!
1 Remove holding screws
Remove the holding screws 37
37
37
37 and the washer mounted on the
screw threads 12
12
12
12 at tower 27
27
27
27 using the 6.2621.100 Allen key.
2 Remove protective film from splash protection
Remove the plastic film glued on both sides of the splash
protection 1
111.
3 Install splash protection
Attach splash protection onto the screw threads at the tower and
fix it with the holding screws and the washer using the
6.2621.100 Allen key.
37
37
37
37
37
37
37
37
1111
Fig. 5: Installing the splash protection
2.3.5
Installing the needle or tubing
1 Remove PEEK compression fitting 5
Remove PEEK compression fitting 5
555 screwed onto the needle
holder 4
444.
2 Insert needle
Insert needle 2
222 (6.1835.010) PEEK needle or 6.1835.020 PEEK
tubing) completely into the opening of the needle holder 4
444.
3 Fix needle or tubing
Tighten compression fitting 5
555 in needle holder 4444 by hand
(never use tools!).
1111 Splash protection
(6.2751.040)
37
37
37
37 Holding screws