Metrohm 788 IC Filtration Sample Processor User Manual
Page 37
2.4 Installation for full control with «IC Net»
788 IC Filtration Sample Processor
29
Settings in the «IC Net 2.1» program
1. Definition of new program parameters for the 788 IC Filtration
Sample Processor under Configuration/Control:
FILL A 2
= ---***0*010***
INJECT A 2
= ---***1*000***
ZERO 2
= ---***0*011***
2. Program for the 788 IC Filtration Sample Processor:
001 Ctrl INIT
002 Move
sample
003 Lift work
004 Ctrl ZERO
1
005
Ctrl
FILL A 1
006
Ctrl
STEP MSM 753
007
Pump 240 s
008
Ctrl
INJECT A 1
009 Move
sample+1
010 Lift work
011 Ctrl ZERO
2
012 Ctrl Init
732
013 Scan
Wait1
014
Ctrl
FILL A 2
015 Ctrl Init
732
016 Pump
240
s
017
Ctrl
INJECT A 2
018 Ctrl Init
732
Initialize remote interface
Move needle to sample position
Place lift with needle to working position
Trigger autozero at 732/1 IC Detector
Switch injection valve A at 733 to "Fill"
Switch 753 suppressor module to next position
Fill sample loop A with sample during 240 s
Switch injection valve A at 733 to "Inject"
Move needle to sample position (+1)
Place lift with needle to working position
Trigger autozero at 732/2 IC Detector
Initialize remote lines at 732/1 and 732/2
Wait until 732/1 IC Det. sends signal on remote line 3
Switch injection valve B at 733 to "Fill"
Initialize remote lines at 732/1 and 732/2
Fill sample loop B with sample during 240 s
Switch injection valve B at 733 to "Inject"
Initialize remote lines at 732/1 and 732/2
3. Definition of new program parameters for the 723/1 IC Detector
under Programm/Remote Configuration:
Start_788
=
**1*****
Reset_Start_788
=
**0*****
4. To synchronize the second injection to the start of the second
chromatogram, a time program is started for the 732/1 IC Detector
directly with the first injection:
1.0 Start_788
1.1
Reset_Start_788
1.2
Flag end
Both detectors 732/1 and 732/2 are registered in the system window
under Setup/Start mode in the field 'Start with inject' .
4. By this procedure, with a delay of one minute, the 788 IC Filtration
Sample Processor receives a signal for processing the second
sample. The total resulting delay (5 min.) is now independent from
the time the sampler needs for rotating to the next sample.
Therefore, the method for the second sample injection (e.g. method
'cation') gets a start delay under Method Setup/Measure:
Start delay:
5.0 min