2 configuration, Stop – Metrohm 788 IC Filtration Sample Processor User Manual
Page 54
3 Operating tutorial
788 IC Filtration Sample Processor
46
3.2 Configuration
CONFIG
ENTER
TOWER
SELECT
x The dialog language can be set in the configuration menu. Press
configuration
>auxiliaries
x and then
>auxiliaries
dialog: english
x This menu item has a colon, indicating that here the parameters can
be selected from a list. Press
view the various selections and get used to this type of dialog.
>auxiliaries
dialog: deutsch
x With
'dialog:
english'
.
>auxiliaries
display contrast: 3
4 u <Ð>
x By pressing <Ð> 4 times you reach the menu selection
'max. lift
way'
.
>auxiliaries
max. lift way 125 mm
x Here the lowest allowable lift position for automatic and manual
operation can be set. For the 6.2041.430 sample rack with
6.2743.050 sample tubes, this limit value of 125 mm should not be
changed. Accept the value by pressing
configuration
>rack definitions
QUIT
or
STOP
x In order to put the 788 IC Filtration Sample Processor back into the
initial position, press
******** counter 0/127
PUMP- ready
x In the normal state, the method name and the sample counter read-
ing are displayed in the first line. The second line serves as status
line which displays the pump status and the changer status.