Metrohm 788 IC Filtration Sample Processor User Manual
Page 59
3.5 "Tracing"
788 IC Filtration Sample Processor
51
HOLD
LEARN
x Press
LED indicates that the 788 IC Filtration Sample Processor is ready to
execute the command.
x Now move the lift into the desired position with the <Ð> and <Ï>
keys. You will notice that the current lift position is always indicated
"live". During execution of the command the LEARN-LED is lit con-
tinuously. Accept the lift position that has been set by pressing
goes off again.
>sample sequence
4 CTL:Rm: FILL A 1
x In this line the injection valve at the 733 IC Separation Center is
switched to the "Fill" position.
HOLD
LEARN
HOLD
LEARN
>sample sequence
5 PUMP 1.1 : 240 s
x In this line the peristaltic pump at the 788 IC Filtration Sample Proc-
essor is started in order to convey the sample from the sample tube
to the sample loop at the 733 IC Separation Center.
x Here, you can use the LEARN mode to optimize the pump time too.
x In this case, as with the other "teachable" commands (the LIFT
command is an exception), pressing the
immediate execution of the corresponding command. The elapsed
time is also displayed here "live". By pressing the
again the command can be interrupted.
x The blinking LED indicates that the 788 IC Filtration Sample Proces-
sor is still in the LEARN mode. If you now switch the pump back on
with the
is now added to the existing value.
x Now set the pump time in this way. Accept the total time with
>sample sequence
6 CTL:Rm: ZERO 1
x In this line the autozero function at the 732 IC Detector is triggered.
>sample sequence
7 CTL:Rm: INJECT A 1
x In this line the injection valve at the 733 IC Separation Center is
switched to the "Inject" position.