6 placing the sample rack, 7 adjusting the sample rack – Metrohm 788 IC Filtration Sample Processor User Manual
Page 22
2 Installation
788 IC Filtration Sample Processor
14
2222
4444
5555
6666
7777
8888
Fig. 6: Needle installation
If you are using the 6.1835.020 PEEK tubing, the sample tubes may
not be sealed with caps because they cannot be pierced by the PEEK
tubing and the needle may be damaged thereby.
2.3.6
Placing the sample rack
1
2
Place sample rack
Place sample rack 35
35
35
35 on the
turntable of the 788 IC Filtration
Sample Processor according
to Fig. 7.
Read magnetic code
Press
the rack to the home position,
in which the magnetic code
can be read (details see
section 4.5).
Fig. 7: Sample rack placing
2.3.7
Adjusting the sample rack
If a new sample rack is placed on the 788 IC Filtration Sample
Processor for the first time, it must be adjusted on a rack position in the
middle row (example: position 37) as follows:
1 Place sample rack
Place empty sample rack on the 788 IC Filtration Sample
Processor and press
2222 PEEK needle (6.1835.010) or
PEEK tubing (6.1835.020)
4444 Needle
holder
5555 PEEK compression fitting
(4.766.4320)
6666 PEEK compression fitting
(6.2744.010)
7777 Swing
head
8888 PTFE capillary (6.1803.070)