788 IC Filtration Sample Processor

142

Required Instruments:

Only required, if the RS232 or remote interface is to be checked:

x 3.496.8550 Test plug (on the "Remote" plug)
x 3.496.8480 Test plug (on the "RS 232" plug)

6.4.2

Preparing the instrument

x Disconnect power plug.

x Remove cables to the "RS232" and "Remote" interfaces.

x Plug in power plug and immediately press <9> and keep it de-

pressed until the start test pattern disappears.

x Alternatively, the key can be pressed to cause a reset fol-

lowed by pressing (within 0.4 sec) and holding down the <9> key to

open the diagnosis menu.

Switch on and

9

LIFT

or

RESET

CLEAR

and

9

LIFT

Main menu Diagnosis:

diagnosis
>RAM initialization

Open the submenu with

diagnosis
>RAM test

Use <Ï> or <Ð> to access one menu

item up or down

diagnosis
>display test

Use or to access first

or last menu item

diagnosis
>display contrast test

returns to the normal state

diagnosis
>key test

diagnosis
>remote test

diagnosis
>RS232 test

diagnosis
>external bus test

diagnosis
>beeper test

diagnosis
>rack code test

diagnosis
>function test

diagnosis
>instrument number

diagnosis
>power on reset

The "function test" is a Metrohm internal test which should be run only

by service personal. Before starting this test, the needle must always

be removed from the needle holder, otherwise it would be damaged.