2 preparing the instrument – Metrohm 788 IC Filtration Sample Processor User Manual
Page 150
6 Appendix
788 IC Filtration Sample Processor
142
Required Instruments:
Only required, if the RS232 or remote interface is to be checked:
x 3.496.8550 Test plug (on the "Remote" plug)
x 3.496.8480 Test plug (on the "RS 232" plug)
6.4.2
Preparing the instrument
x Disconnect power plug.
x Remove cables to the "RS232" and "Remote" interfaces.
x Plug in power plug and immediately press <9> and keep it de-
pressed until the start test pattern disappears.
x Alternatively, the
lowed by pressing (within 0.4 sec) and holding down the <9> key to
open the diagnosis menu.
Switch on and
9
LIFT
or
RESET
CLEAR
and
9
LIFT
Main menu Diagnosis:
diagnosis
>RAM initialization
Open the submenu with
diagnosis
>RAM test
Use <Ï> or <Ð> to access one menu
item up or down
diagnosis
>display test
Use
or last menu item
diagnosis
>display contrast test
diagnosis
>key test
diagnosis
>remote test
diagnosis
>RS232 test
diagnosis
>external bus test
diagnosis
>beeper test
diagnosis
>rack code test
diagnosis
>function test
diagnosis
>instrument number
diagnosis
>power on reset
The "function test" is a Metrohm internal test which should be run only
by service personal. Before starting this test, the needle must always
be removed from the needle holder, otherwise it would be damaged.