26 • Start Up
EAGLE 2 Operator’s Manual
will again appear in the startup sequence.
•
If the unit is due for calibration and CAL PAST DUE ACT is set to
NOTIFICATION ONLY, then the following alert screen displays and
the buzzer sounds in a double pulsing pattern.
Press and release the POWER ENTER RESET button to
acknowledge the alert and continue with the startup sequence.
10. The Date/Time Screen appears for a few seconds.
11. If USER/STATION ID is turned on (see “Turning the User/Station ID
Function On or Off” on page 118), the ID Screen appears for a few
seconds.
If USER/STATION ID is turned off, only the serial number is shown.
12. If the EAGLE 2 experiences a sensor failure during start up, a screen
indicating which sensor failed appears and the buzzer sounds a pulsing
tone twice per second. In the example below, the H
2
S sensor has failed.
CALIBRATION DATE
IS PAST DUE
9/12/2008
15:00:00
USER ID
MIKE
STATION ID
PUMP 1
SERIAL NUMBER
E2A515
FAILED SENSOR(S)
< > < >
ENTER TO CONTINUE