176 • Configuring the PID Gas in Setup Mode
EAGLE 2 Operator’s Manual
13. Press and release the POWER ENTER RESET button. The LCD will ask
for confirmation that you want to change the configured gas.
14. To change the gas configuration, press and release AIR ▲ YES. The unit
will return to the Configure Gases Screen reflecting the new gas
configuration.
If you do not want to change the gas configuration, press and release
DISPLAY ADJUST NO. The unit will return to the first gas
configuration choice screen with the gas configuration unchanged.
Select a different gas and go to Step 13 or press and release DISPLAY
ADJUST NO to return to the Gas Configuration screen.
15. Use RANGE ▼ SHIFT to place the cursor next to END.
16. Press and release POWER ENTER RESET to return to the main menu.
17. Use RANGE ▼ SHIFT to place the cursor in front of NORMAL
OPERATION at the bottom of the main menu.
18. Press and release POWER ENTER RESET.
CHANGE TO IPA ?
PRESS YES OR NO
CONFIGURE GASES
CAT : CH4 (CAT)
> OP1 : IPA (PID)
OP2 : --- (---)
OP3 : --- (---)
PID(0- 2000ppm IBL )
> 1000 ppm ACT
1000 ppm BNZ
1500 ppm DSL
15000 ppm ETA