Start up and normal operation – RKI Instruments EAGLE 2 Manual User Manual
Page 148

138 • Start Up and Normal Operation
EAGLE 2 Operator’s Manual
use the probe intended for PID use that does not have a particle filter installed
and that you install the probe directly to the inlet fitting when monitoring for
gas with the PID sensor. The figure below shows an exploded drawing of the
PID probe.
Figure 26: PID Probe
Start Up and Normal Operation
For instructions to startup and use an EAGLE 2 that includes a PID sensor,
reference “Start Up” on page 22, “Measuring Mode, Normal Operation” on
page 29, and “Measuring Mode, Alarms” on page 35. Follow these
instructions keeping the following special considerations in mind:
•
After the instrument’s warm up period is over and the gas readings are
displayed on the screen, if the PID sensor is still in its warm up state, the
gas reading for the PID channel will scroll through -200, -100, and 0 for a
high range sensor and -2.00, -1.00, and 0.00 for a low range sensor. Allow
the sensor to continue warming up for 5 minutes. If it exits its warm up