Start up – RKI Instruments EAGLE 2 Manual User Manual
Page 306

296 • Start Up
EAGLE 2 Operator’s Manual
13. When you are done making changes, use RANGE ▼ SHIFT to move the
cursor next to END.
14. Press and release POWER ENTER RESET to save the new settings and
return to the Change Alarm Point Settings Screen.
15. Use RANGE ▼ SHIFT to move the cursor next to END.
16. Press and release POWER ENTER RESET to return to the main menu.
Start Up
1. Press and briefly hold down the POWER ENTER RESET button.
Release the button when you hear a beep.
2. The LCD will show the following screen for about ten seconds.
3. The Battery Voltage Screen appears for a few seconds.
4. The Inert Mode Selection Screen will then be displayed.
Use the AIR ▲ YES or RANGE
▼
SHIFT button to display the mode you
wish to enter. If you have not made a selection after 20 seconds, the
instrument will begin to beep and the LEDs will begin to flash. They will
continue until you make a selection.
GAS MONITOR
EAGLE 2
BATTERY MIN:
4.3 VOLTS
BATTERY NOW:
5.2 VOLTS
MODE SELECT
UP/DOWN THEN ENTER
NORMAL MODE