RKI Instruments EAGLE 2 Manual User Manual
Page 207

EAGLE 2 Operator’s Manual
ESM-01 Calibration • 197
29. If you want to change the ESM-01 channel’s calibration gas value, follow
Step 15 - Step 24 above beginning with the ESM-01 Calibration Gas
Value Screen below instead of the standard channel Calibration Gas
Value Screen.
30. With the ESM-01 Calibration Gas Value Screen displayed, press and
release the POWER ENTER RESET button to proceed to the Calibration
In Process Screen for the ESM-01 channel with CAL IN PROCESS
flashing.
If you do not want to proceed with the calibration, press and release the
DISPLAY ADJUST NO button to return to the ESM-01 Cal Gas Values
Screen.
If you do want to continue with the calibration, proceed to the next step.
31. Since calibration gas has already been flowing to the ESM-01 sensor
while the other channels were being calibrated, the ESM-01 sensor
reading should be stable and ready for adjustment.
32. Press and release the POWER ENTER RESET button to set the span
adjustment for the ESM-01 channel to the programmed value.
CAL GAS VALUE
SO2 5 ppm
ENTER TO BEGIN CAL
CAL IN PROCESS
SO2 3 ppm
ENTER WHEN DONE