106 • Using Setup Mode
EAGLE 2 Operator’s Manual
configuration.
CAUTION: Before changing the channel configuration, confirm that the
correct sensors and electronic hardware are installed in the
EAGLE 2 and that its construction and flow system are
appropriate for the installed sensors. Operation of the EAGLE 2
with a flow system or construction not compatible with the
installed sensors will result in inaccurate readings. Consult RKI
Instruments, Inc. if you cannot confirm either of these items.
1. From the main menu, place the cursor next to CONFIGURE
CHANNELS.
2. Press and release POWER ENTER RESET. The Configure Channels
screen appears with the cursor flashing next to CH1.
The standard 4-gas configuration is shown below.
3. Use AIR ▲ YES or RANGE ▼ SHIFT to move the cursor next to the
channel you want to configure.
CONFIGURE CHANNELS
> CH1: CAT (CH4)
CH4: CO (CO)
CH2: OXY (OXY)
CH3: H2S (H2S)
CONFIGURE CHANNELS
CH1: CAT (CH4)
> CH4: CO (CO)
CONFIGURE CHANNELS
> CH5: NOT USED
AIR
YES
RANGE
SHIFT
CH2: OXY (OXY)
CH3: H2S (H2S)
CH6: NOT USED
END