Parts list, Channel as described in “infrared co, Calibration” on page 227 – RKI Instruments EAGLE 2 Manual User Manual
Page 247

EAGLE 2 Operator’s Manual
Parts List • 237
properly, it may be damaged when the top case is re-installed.
16. Make sure that the top case gasket is fully seated in its groove and
carefully put the top case back on the bottom case. If you have any
difficulty mating the top and bottom cases, inspect the placement of the
main PCB and the placement of the top case gasket.
17. Turn the EAGLE 2 upside down and tighten the three case screws to
secure the top case to the bottom case.
18. Calibrate the infrared CO
2
channel as described in “Infrared CO
2
Calibration” on page 227.
Parts List
Table 25: Infrared CO
2
EAGLE 2 Parts List
Part Number
Description
47-5051RK
Infrared CO
2
sensor cable
DEM-3313-1
Infrared CO
2
sensor, 0-5 %vol
DEM-3313-4
Infrared CO
2
sensor, 0-10,000 ppm
DEM-3313-5
Infrared CO
2
sensor, 0-60 %vol
81-0071RK-01
Calibration cylinder, 5000 ppm CO
2
in nitrogen, 34 liter
81-0071RK-03
Calibration cylinder, 5000 ppm CO
2
in nitrogen, 103 liter
81-0072RK-01
Calibration cylinder, 2.5 %vol CO
2
in nitrogen, 34 liter
81-0072RK-03
Calibration cylinder, 2.5 %vol CO
2
in nitrogen, 103 liter
81-0073RK-01
Calibration cylinder, 15 %vol CO
2
in nitrogen, 34 liter
81-0073RK-03
Calibration cylinder, 15 %vol CO
2
in nitrogen, 103 liter