Calibrating using the single calibration method – RKI Instruments EAGLE 2 Manual User Manual
Page 104

94 • Calibrating Using the Single Calibration Method
EAGLE 2 Operator’s Manual
If any of the sensors cannot be adjusted to the proper value, a screen
displays that indicates a calibration failure and lists the sensor(s) that
failed to calibrate. In the example below, the oxygen and H
2
S channels
failed calibration. The other sensors calibrated normally.
The buzzer and LED arrays activate in a double pulsing pattern. Press
and release the POWER ENTER RESET button to reset the alarm and
return to the Calibration Mode Screen. Attempt to calibrate again. If the
failure continues, investigate the cause. See “Troubleshooting” on
page 70
25. Disconnect the tubing from the probe.
26. Use the RANGE ▼ SHIFT button to place the cursor next to the
NORMAL OPERATION menu option, then press and release the
POWER ENTER RESET button to return to Measuring Mode.
Calibrating Using the Single
Calibration Method
Single Calibration allows you to calibrate one channel at a time. This is
useful if you only want to calibrate one or two channels.
Setting the Fresh Air Reading
1. While in the Calibration Mode Screen, move the cursor to the
PERFORM AIR ADJUST menu item by using the RANGE ▼ SHIFT
button.
FAIL SENSOR
< >
ENTER TO CONTINUE
CALIBRATION MODE
AUTO CALIBRATION
SINGLE CALIBRATION
> PERFORM AIR ADJUST
NORMAL OPERATION