268 • Monitoring in Methane Elimination Mode
EAGLE 2 Operator’s Manual
WARNING: The EAGLE 2’s alarms are initiated by the display reading,
not the factored reading. If you are monitoring for isobutane
as in the above example and the low alarm is set for 10%
LEL, the EAGLE 2 will initiate a low alarm at 5.9% LEL
isobutane (display reading of 10% LEL).
12. To return to full response mode, return to the Methane Elimination Mode
screen in Display Mode and press and release the AIR ▲ YES or
RANGE
▼
SHIFT button to toggle the setting to DISABLED.
13. Press and release the POWER ENTER RESET button. The unit will save
the setting and proceed to the next menu item.
14. Continue to press the DISPLAY button until the Measuring Mode screen
appears.
The ME is no longer next to the catalytic combustible channel reading
1-Butyl Alcohol
1.71
2.18
Methyl Acrylate
0.83
2.11
Cyclohexane
1.06
1.25
Methyl Ethyl
Keytone
1.04
1.32
Cumene
1.74
1.42
Methyl Isobutyl
Keytone
1.30
1.42
Ethylene
Dichloride
2.04
11.50
Mixed Xylenes
1.36
1.36
Ethyl Alcohol
0.57
1.71
Nonane
1.66
1.21
Ethyl Chloride
0.59
2.04
Pentane
0.54
0.74
Ethyl Acrylate
1.32
1.68
Propane
Low
response
Low
response
Hexane
1.00
1.00
Styrene
1.74
1.42
Hydrogen
0.48
1.75
Toluene
1.25
1.25
Vinyl Acetate
Monomer
0.92
2.17
*
Vapor pressure too low for significant LEL reading
Target Gas
LEL Factor
PPM Factor
Target Gas
LEL Factor
PPM Factor
HEX 0%LEL
OXY 20.9vol%
H2S 0.0ppm
CO 0ppm