RKI Instruments EAGLE 2 Manual User Manual
Page 134

124 • Using Setup Mode
EAGLE 2 Operator’s Manual
2. Press and release POWER ENTER RESET. The Calibration Past Due
Action Screen appears.
3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.
4. Press and release POWER ENTER RESET to save the setting and return
to the main menu.
Updating the Calibration Interval
This setting defines the amount of time between calibrations. The time can be
set in 1 day increments. The minimum setting is 1 day and the maximum
setting is 365 days. The factory setting is 90 days.
1. From the main menu, place the cursor next to CAL INTERVAL.
2. Press and release POWER ENTER RESET. The Set Calibration Interval
Screen appears.
3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.
Tip: Press and hold AIR ▲ YES or RANGE ▼ SHIFT to rapidly scroll
through the settings.
4. Press and release POWER ENTER RESET to save the setting and return
to the main menu.
Updating LC/BH Mode Setting
With LC/BH MODE SELECT set to LC & BH, the Mode Select Screen
appears when the unit is turned on. You are able to select from Normal Mode,
Leak Check Mode, and Bar Hole Mode.
With LC/BH MODE SELECT set to BAR HOLE ONLY, the Mode Select
Screen appears when the unit is turned on. You are able to select from
Normal Mode and Bar Hole Mode.
CALIBRATION PAST
DUE ACTION
UP/DOWN THEN ENTER
CONFIRM TO CAL
SET CALIBRATION
INTERVAL
UP/DOWN THEN ENTER
90 DAYS