RKI Instruments EAGLE 2 Manual User Manual
Page 183

EAGLE 2 Operator’s Manual
Configuring the PID Gas in Setup Mode • 173
4. The “S” will then disappear and the following screen will appear for a
few seconds.
5. If the unit prompts you for the password, enter it by using AIR ▲ YES
and RANGE ▼ SHIFT to select each password number and then pressing
and releasing the POWER ENTER RESET button to enter it and move on
to the next number until all of the numbers are entered. The main menu
displays. It displays six menu items at a time.
6. Use the RANGE ▼ SHIFT button to move the cursor down the menu to
CONFIGURE GASES.
7. Press and release POWER ENTER RESET. The Configure Gases Screen
appears with the cursor flashing next to CAT, the catalytic sensor. If an
EAGLE 2 has a PID sensor installed, one of the three optional sensor
types, OP1, OP2, or OP3 will indicate it is a PID. In the example below,
OP1 is shown as a PID sensor.
GAS MONITOR
EAGLE 2
>SET DATE & TIME
SET DATE FORMAT
SET BATTERY TYPE
CONFIGURE CHANNELS
CONFIGURE GASES
CATALYTIC UNITS
CONFIGURE GASES
> CAT : CH4 (CAT)
OP1 : IBL (PID)
OP2 : --- (---)
OP3 : --- (---)