RKI Instruments EAGLE 2 Manual User Manual
Page 151

EAGLE 2 Operator’s Manual
PID Relative Response Feature • 141
2. With the PID Sensor Relative Response Screen displayed, press and
release AIR ▲ YES. A list of gases will appear on the screen with EXIT
at the top of the list.
There are multiple screens of gases. The following is the complete list of
factory defined gases along with their detection ranges, low alarm, high
alarm, STEL, and TWA settings. Table 14 is a list of the low range values
and Table 15 is a list of the high range values.
Table 14: Low Range Relative Response Gas List
Target Gas
Detection
Range (ppm)
Alarm 1 Factory
Setting (ppm)
Alarm 2 Factory
Setting (ppm)
STEL
(ppm)
TWA
(ppm)
Acetone
0-30.00
5.00
7.50
OFF
OFF
Benzene
0-25.00
0.50
2.50
2.50
0.50
Diesel Fuel
NO 1
0-40.00
2.00
3.00
OFF
OFF
Ethanol
0-400.00
10.0
15.0
OFF
OFF
Gasoline
0-50.00
3.00
5.00
OFF
OFF
Isobutylene
0-50.00
4.30
6.00
OFF
OFF
Isopropanol
0-200.00
2.0
4.0
OFF
OFF
JP-5 Fuel
0-30.00
1.40
2.10
OFF
OFF
Methyl Ethyl
Ketone
0-40.00
2.00
3.00
OFF
OFF
Toluene
0-25.00
0.50
1.50
OFF
OFF
N-Hexane
0-200.00
5.0
10.0
OFF
OFF
Propylene
0-50.00
5.00
7.50
OFF
OFF
Styrene
0-20.00
0.20
0.40
OFF
OFF
Tetrachloro-
ethylene
0-30.00
0.24
1.00
OFF
OFF
Trichloro-
ethylene
0-30.00
0.50
1.00
OFF
OFF
Vinyl Chloride
0-100.00
1.0
5.0
5.0
1.0
>EXIT
ACETONE
BENZENE
DIESEL FUEL NO 1
ETHANOL
GASOLINE