RKI Instruments EAGLE 2 Manual User Manual
Page 184

174 • Configuring the PID Gas in Setup Mode
EAGLE 2 Operator’s Manual
8. Use RANGE ▼ SHIFT to move the cursor down the menu to the PID
sensor.
9. To change the PID sensor gas configuration, press and release POWER
ENTER RESET.
10. A screen appears that indicates the PID type, low range or high range
(flashing), and the detection range for the currently configured gas. In the
example below, the PID sensor is currently configured as a high range
sensor for 0 - 2,000 ppm isobutylene.
11. Use AIR ▲ YES and RANGE ▼ SHIFT to set the correct range, low or
high, for the sensor that is installed.
Press and release POWER ENTER RESET. A screen appears with gas
configuration choices for the PID sensor. The screen below illustrates the
first four choices for a high range PID.
There are 17 choices that can be viewed on five screens. Table 16 below
lists all of the gas configuration choices.
CONFIGURE GASES
CAT : CH4 (CAT)
> OP1 : IBL (PID)
OP2 : --- (---)
OP3 : --- (---)
PID(0- 2000ppm IBL )
UP/DOWN THEN ENTER
HIGH RANGE
PID(0- 2000ppm IBL )
> 1000 ppm ACT
1000 ppm BNZ
1500 ppm DSL
15000 ppm ETA