Configuring the tc gas in setup mode – RKI Instruments EAGLE 2 Manual User Manual
Page 227

EAGLE 2 Operator’s Manual
Configuring the TC Gas in Setup Mode • 217
properly, it may be damaged when the top case is reinstalled.
16. Make sure that the top case gasket is fully seated in its groove and
carefully put the top case back on the bottom case. If you have any
difficulty mating the top and bottom cases, inspect the placement of the
main PCB and the placement of the top case gasket.
17. Turn the EAGLE 2 upside down and tighten the three case screws to
secure the top case to the bottom case.
18. Calibrate the TC channel as described in “TC Calibration” on page 207.
Configuring the TC Gas in
Setup Mode
The TC channel can be configured for 3 different pre-defined gases and 1
user-defined gas in the CONFIGURE GASES menu item in Setup Mode. To
change the gas configuration of the TC channel in Setup Mode, do the
following:
WARNING: Do not use the user defined gas configuration without
consulting RKI Instruments, Inc.
WARNING: The EAGLE 2 is not in operation as a gas detector while in
Setup Mode.
1. Take the EAGLE 2 to a non-hazardous location and turn it off if it is on.
2. Press and hold AIR ▲ YES and RANGE ▼ SHIFT, then press and hold
POWER ENTER RESET. When you hear a beep, release the buttons.
3. The LCD will show the following screen for a few seconds with the “S”
in the lower right corner indicating the unit is entering Setup Mode.
GAS MONITOR
EAGLE 2 S