Campbell Scientific CR9000X Measurement and Control System User Manual
Page 419

Appendix D. CR9000X Status Table
Field Name
Variable
Type
Description
BuffDepth
Integer
Shows the processing buffer depth (# of scans that processing is lagging the
measurements). Indicates how far processing is currently behind measurement.
MaxBuffDepth
Integer
Gives the maximum number of buffers processing lagged measurement.
Indicative of how close the program is to skipping scans.
LastSlowScan Time Time
of
the last Slow Sequence Scan.
SlowProcTime
Integer
Processing time (μSeconds) of the last Slow Sequence scan. Processing occurs
concurrent with measurement so the sum of measure time and process time is
not the time required in the scan instruction.
MaxSlowProcTime Integer Maximum process time (
μSeconds) required, as yet, for the processing of the
measurement values from one scan of the Slow Sequence Table.
CalVolts Integer
Array
Factory calibration numbers. This array contains six values corresponding to
the six measurement integration options as shown in the following table. These
numbers are loaded during the Factory Calibration and are stored in FLASH.
# Value
Voltage Range
1
5000
mV
2
1000
mV
3
500
mV
4
500
mVX
5
200
mV
6
50
mV
CalGain(#) Integer
Array
Displays the Gain calibration factor for the different voltage ranges.
CalGain(#) shows the calibration factor for the voltage ranges as depicted in
the following table. These values are updated at program compile time or
when a Calibrate or BiasComp instruction is encountered in the program, if the
program uses the measurement range.
# Value
Voltage Range
1
5000
mV
2
1000
mV
3
500
mV
4
500
mVX
5
200
mV
6
50
mV
CalOffSet(#) Integer
Array
Displays the Offset calibration factor for the different voltage ranges.
CalOffset(#) shows the calibration factor for the voltage ranges as depicted in
the following table. These values are updated at program compile time or when
a Calibrate or BiasComp instruction is encountered in the program, if the
program uses the measurement range.
# Value
Voltage Range
1
5000
mV
2
1000
mV
3
500
mV
4
500
mVX
5
200
mV
6
50
mV
D-3