Voltfilt parameters – Campbell Scientific CR9000X Measurement and Control System User Manual
Page 269

Section 7. Measurement Instructions
7-45
Parameter
Enter
VOLTFILT PARAMETERS
Dest
Variable,Array
The Variable to store the results of the instruction. When Reps are used the results are stored in an
array with the variable name. An array must be dimensioned to have elements for all Reps.
Reps
Constant
The number of times to repeat the measurement on subsequent CR9052 channels.
.
Range
The voltage range for the measurement. The CR9052 normally replaces out-of-range
measurements with not-a-number (NaN) which is displayed in RTDaq's real time windows as
Range?
. Users may choose to have out-of-range measurements to be replaced by the analog-to-
digital converter saturation value with a special code in FiltOption.
Constant
Alpha
Code
Numeric
Code
Voltage
Range
Module Excitation
Board Supported
mV5000
0
± 5000 mV
CR9052DC, CR9052IEPE
mV1000
1
± 1000 mV
CR9052DC, CR9052IEPE
mV200
4
± 200 mV
CR9052DC
mV50
5
± 50 mV
CR9052DC
mV20
6
± 20 mV
CR9052DC
Fslot
Constant
The number of the slot that holds the CR9052 Module to be used for the measurement.
Chan
Constant
The CR9052 channel number on which to make the first measurement. When Reps are used,
subsequent measurements will be automatically made on the following differential channels.
FiltOption
Constant
The sample ratio for the measurement (how many measurements are made within one cycle of the
highest frequency in the pass band). The sample ratio determines the top of the pass-band (F
PASS
)
and the beginning of the stop-band (F
STOP
) of the anti-aliasing low-pass filter relative the sample
rate (F
SAMPLE
). The sample rate is the inverse of the scan interval in the CRBASIC program.
FiltOption
must be the same for all channels of a single CR9052 Filter Module. Out-of-range
measurements may be replaced by the analog-to-digital converter saturation value by adding 1000
to the FiltOption codes shown below.
NumericCode Sampling
Ratio
F
PASS
F
STOP
2 2.5
F
SAMPLE
/2.5
F
SAMPLE
/2.01
5 5
F
SAMPLE
/5
F
SAMPLE
/3.37
10 10
F
SAMPLE
/10
F
SAMPLE
/5.08
20 20
F
SAMPLE
/20
F
SAMPLE
/6.81
1* 2.155
23.2 Khz
26.8 kHz
*
Option 1 has no additional filtering beyond the CR9052DC analog front-end and the sigma-delta
A/D converter, thus freeing the CR9052DC on-board digital signal processor for additional
processing. F
SAMPLE
must be 50 kHz to use this filter option.
Excitation
Constant
The continuous, dc, output for the excitation channel(s). If Reps is greater than one, then the same
excitation will be output on sequential excitation channels.
Numeric Code
Alpha Code
Output Level
IEPE Freq. Response
CR9052IEPE
900 (905)
None
None
τ
= 0.5 Sec (
τ
= 5.0 Sec)
CR9052IEPE
605
None
Constant 6 mA
0.3 Hz to 20 kHz (
τ
= 5.0 Sec)
CR9052IEPE
405
None
Constant 4 mA
0.3 Hz to 20 kHz (
τ
= 5.0 Sec)
CR9052IEPE
205
None
Constant 2 mA
0.3 Hz to 20 kHz (
τ
= 5.0 Sec)
CR9052IEPE
600
None
Constant 6 mA
3.0 Hz to 20 kHz (
τ
= 0.5 Sec)
CR9052IEPE
400
None
Constant 4 mA
3.0 Hz to 20 kHz (
τ
= 0.5 Sec)
CR9052IEPE
200
None
Constant 2 mA
3.0 Hz to 20 kHz (
τ
= 0.5 Sec)
CR9052DC
7
V10
Constant 10 V DC
N/A
CR9052DC
5
None
Constant 5 V DC
N/A
CR9052DC
2
None
Constant 10 mA
N/A
CR9052DC
1
None
None N/A
Mult, Offset
Constant,Variable,
Array, Expression
A multiplier and offset by which to scale the measurement. A multiplier of one and an offset of 0
will return the measurement in millivolts.