beautypg.com

Accuracy of the thermocouple voltage measurement – Campbell Scientific CR9000X Measurement and Control System User Manual

Page 101

background image

Section 3. CR9000X Measurement Details

TABLE 3.1.4-1. Limits of Error for Thermocouple Wire (Reference

Junction at 0

o

C)

Limits of Error

Thermocouple Temperature

(Whichever is greater)

Type Range

o

C Standard Special

T

-200

to

0

±

1.0

o

C or 1.50%

0

to

350

±

1.0

o

C or 0.75%

±

0.5

o

C or 0.4%

J

0

to

750

±

2.2

o

C or 0.75%

±

1.1

o

C or 0.4%

E

-200

to

0

0

to

900

±

1.7

o

C or 1.00%

±

1.7

o

C or 0.50%

±

1.0

o

C or 0.4%

K

-200

to

0

0

to

1250

±

2.2

o

C or 2.00%

±

2.2

o

C or 0.75%

±

1.1

o

C or 0.4%

N

-270

to

0

0

to

1300

±

2.2

o

C or 2.00%

±

2.2

o

C or 0.75%

±

1.1

o

C or 0.4%

R or S

0 to 1450

±

1.5

o

C or 0.25%

±

0.6

o

C or 0.1%

B

800

to

1700

±

0.5%

Not Estab.


When both junctions of a thermocouple are at the same temperature there is no
voltage produced (law of intermediate metals). A consequence of this is that a
thermocouple can not have an offset error; any deviation from a standard
(assuming the wires are each homogeneous and no secondary junctions exist)
is due to a deviation in slope. In light of this, the fixed temperature limits of
error (e.g.,

±

1.0

°C for type T as opposed to the slope error of 0.75% of the

temperature) in the table above are probably greater than one would experience
when considering temperatures in the environmental range (i.e., the reference
junction, at 0

°C, is relatively close to the temperature being measured, so the

absolute error - the product of the temperature difference and the slope error -
should be closer to the percentage error than the fixed error). Likewise,
because thermocouple calibration error is a slope error, accuracy can be
increased when the reference junction temperature is close to the measurement
temperature. For the same reason differential temperature measurements, over
a small temperature gradient, can be extremely accurate.

In order to quantitatively evaluate thermocouple error when the reference
junction is not fixed at 0

o

C, one needs limits of error for the Seebeck

coefficient (slope of thermocouple voltage vs. temperature curve) for the
various thermocouples. Lacking this information, a reasonable approach is to
apply the percentage errors, with perhaps 0.25% added on, to the difference in
temperature being measured by the thermocouple.

Accuracy of the Thermocouple Voltage Measurement

The accuracy of a CR9000X voltage measurement is specified as 0.07% the
measured voltage plus 4 A/D counts of the range being used to make the
measurement. The input offset error reduces to 1 A/D count if a differential
measurement is made utilizing the option to reverse the differential input.

For optimum resolution, the

±50 mV range is used for all but high

temperature measurements (Table 3.1.4-2). The input offset error dominates
the voltage measurement error for environmental measurements. A
temperature difference of 40 to 60 °C between the measurement and reference

3-13