2 thermocouple measurements, Voltse parameters – Campbell Scientific CR9000X Measurement and Control System User Manual
Page 229

Section 7. Measurement Instructions
7-5
Parameter
& Data Type
Enter
VOLTSE PARAMETERS
Dest
Var. or Array
The Variable in which to store the results of the instruction. When Reps are used the results are stored in
an array with the variable name. An array must be dimensioned to have elements for all the Reps.
Reps
Constant
The number of repetitions for the measurement.
Range
The voltage range for the measurement. V ranges output volts, mV ranges output millivolts.
Constant
± 5 Volt Analog Input Module
± 50 Volt Analog Input Module
Alpha
Code
Numeric
Code
Voltage
Range
Alpha
Code
Numeric
Code
Voltage
Range
mV5000
0
± 5000 mV
V50
6
± 50 V
mV1000
1
± 1000 mV
V10
7
± 10 V
mV200
4
± 200 mV
V2
10
± 2 V
mV50
5
± 50 mV
mV500
11
± 500 mV
Aslot
Constant
The number of the slot that holds the Analog Input Module to be used for the measurement.
SEChan
Constant
The single-ended channel number on which to make the first measurement. When Reps are used,
subsequent measurements will be made on the sequential single-ended channels. Enter a negative
number to dwell on that channel for the number of measurements specified by the Reps parameter
(except for CR9058E).
SettlingTime
Constant
The time in microseconds to delay between setting up a measurement (switching to the channel, setting
the excitation) and making the measurement (10 microsecond resolution). See Section 3.1.3 Signal
Settling Time.
Entry
Voltage Range
Delay
CR9055
Voltage Range
Delay
0
± 50 mV
20 µS (default)
± 0.5V
40 µS (default)
0
± 200 mV
20 µS (default)
± 2 V
40 µS (default)
0
± 1000 mV
10 µS (default)
± 10 V
30 µS (default)
0
± 5000 mV
10 µS (default)
± 50 V
30 µS (default)
> 0
all
Truncate to closest 10 µS
all
Truncate to closest 10 µS
Integ
Constant
The integration time in microseconds for each of the channels measured (10 microseconds resolution).
See Section 3.1.1.3 Integration for more information on Integration.
Mult, Offset
Constant, Var.,
Array,
Expression
A multiplier and offset by which to scale the raw results of the measurement. See the measurement
description for the units of the raw result; a multiplier of one and an offset of 0 are necessary to output in
the raw units.
7.2 Thermocouple Measurements
TCDiff (Dest, Reps, Range, ASlot, DiffChan, TCType, TRef, RevDiff,
SettlingTime, Integ, Mult, Offset)
Diff. Chanel H
Diff. Chanel L
Thermocouple
This instruction measures a thermocouple with a differential voltage
measurement and calculates the thermocouple temperature (°C) for the
thermocouple type selected. The instruction adds the measured voltage to the
voltage calculated for the reference temperature relative to 0° C, and converts
the combined voltage to temperature in °C. The mV50C and mV200C ranges
briefly (10 µs) connect the differential input to reference voltages prior to
making the voltage measurement to insure that it is within the Input Voltage
Limit range and to test for an open thermocouple.