Section 7. Measurement Instructions
7-8
Parameter
Enter
TCSE PARAMETERS
Dest
Var. or Array
The Variable in which to store the results of the instruction. When Reps are used the results are stored in
an array with the variable name. An array must be dimensioned to have elements for all the Reps.
Reps
Constant
The number of repetitions for the measurement.
*Range
The voltage range for the measurement. TCSE raw output is in Celsius.
Constant
± 5 Volt Analog Input Module
± 50 Volt Analog Input Module
Alpha
Code
Numeric
Code
Voltage
Range
Alpha
Code
Numeric
Code
Voltage
Range
mV5000
0
± 5000 mV
V50
6
± 50 V
mV1000
1
± 1000 mV
V10
7
± 10 V
mV200
4
± 200 mV
V2
10
± 2 V
mV50
5
± 50 mV
mV500
11
± 500 mV
Aslot
Constant
The number of the slot that holds the Analog Input Module to be used for the measurement.
SEChan
Constant
The single-ended channel number on which to make the first measurement. When Reps are used,
subsequent measurements will be made on sequential single-ended channels. Enter a negative number to
dwell on that channel for the number of measurements specified by the Reps parameter (except for
CR9058E).
TCType
The code for the thermocouple type.
Constant
Alpha Code
Numeric Code
Thermocouple Type
TypeT
0 Copper
Constantan
TypeE
1 Chromel
Constantan
TypeK
2 Chromel
Alumel
TypeJ
3 Iron
Constantan
TypeB
4 Platinum
Rhodium
TypeR
5 Platinum
Rhodium
TypeS
6 Platinum
Rhodium
TypeN
7 Nicrosil-Nisil
(NiCRSi-NiSiMg)
TRef
Variable
The name of the variable that is the reference temperature for the thermocouple measurements.
SettlingTime
Constant
The time in microseconds to delay between setting up a measurement (switching to the channel, setting
the excitation) and making the measurement (10 microsecond resolution)
See Section 3.1.3 Signal Settling Time.
Entry
Voltage Range
Delay
CR9055
Voltage Range
Delay
0
± 50 mV
20 µS (default)
± 0.5V
40 µS (default)
0
± 200 mV
20 µS (default)
± 2 V
40 µS (default)
0
± 1000 mV
10 µS (default)
± 10 V
30 µS (default)
0
± 5000 mV
10 µS (default)
± 50 V
30 µS (default)
> 0
all
Truncate to closest 10 µS
all
Truncate to closest 10 µS
Integ
Constant
The integration time in microseconds for each of the channels measured (10 microseconds resolution).
See Section 3.1.1.3 Integration for more information on Integration.
Mult, Offset
Constant, Var,
Array, or
Expression
A multiplier and offset by which to scale the raw results of the measurement. The raw result of the
TCDiff instruction is in degrees C. A multiplier of 1.8 and an offset of 32 will convert the temperature
to degrees F.
*Range: See notes in TCDiff section.
See Section 3.1.4 Thermocouple Measurements for an in-depth study of TC
measurements and an error analysis for them.
See Section 3.1.2.1 Single Ended Voltage Measurements for in-depth
coverage of the Single Ended Measurement process.