Page 240: Am25t parameters

Section 7. Measurement Instructions
7-16
Each element in the array corresponds to a line number in the program. To
accommodate all of the instructions in the program, the array must be
dimensioned to a number greater than or equal to the total number of lines in
the program, including blank lines and comments. The Dest array must also be
dimensioned as a long integer (e.g., Public Array(20) AS LONG).
The execution time for an instruction may vary. For instance, it
may take longer to execute instructions when the datalogger is
communicating with another device.
7.5 Peripheral Devices
AM25T (Dest, Reps, Range, AM25TChan, ASlot, DiffChan, TCType, Tref,
ExCardSlot, ClkPort, ResPort, ExChan, RevDiff, SettlingTime, Integ,
Mult, Offset)
This Instruction controls the AM25T Multiplexer.
Parameter
Enter
AM25T PARAMETERS
Dest
Var. or Array
The Variable in which to store the results of the instruction. When Reps are used the results are stored in
an array with the variable name. An array must be dimensioned to have elements for all the Reps.
Reps
Constant
The number of repetitions for the measurement. For analog measurements, entering reps as a negative
number forces all reps to be on the same channel except for with CR9058E module.
Note: Enter 0 to only measure the output from the PRT.
Range
The voltage range for the measurement. V ranges output volts, mV ranges output millivolts.
Constant
± 5 Volt Analog Input
Module
± 50 Volt Analog Input
Module
Alpha
Code
Num
Code
R *
Option
Code
Voltage
Range
(±mV)
Alpha
Code
Num
Code
R *
Option
Code
Voltage
Range
±
mV5000
0 100
5000 V50 6
N/A
50 V
mV1000
1 101
1000 V10
7
N/A
10 V
mV200
4 104
200 V2
10 N/A
2 V
mV50
5
105
50
mV500 11 N/A
500 mV
mV200C 16
116
200
Alpha Codes ending with a C signify that the channel will be pulled into
See Section
3.2.2 for more
info on the C
& R range
code options.
mV50C
17
117
50
Operational Input Voltage Limits & checked for open input. See Section 3.1.2.
AM25TChan
Constant
The input channel on the AM25T for the first measurement
ASlot
Constant
The number of the slot that contains the CR9050 Module used to measure the AM25T reference
temperature and connected sensors.
DiffChan
Constant
The channel number on the CR9050 Module that will be used to make the actual measurements from the
AM25T.
TCType
The code for the thermocouple type.
Constant
Alpha Code
Numeric Code
Thermocouple Type
mV
-1
Output mV
TypeT
0
Copper Constantan
TypeE
1
Chromel
Constantan
TypeK
2
Chromel
Alumel
TypeJ
3
Iron
Constantan
TypeB
4
Platinum
Rhodium
TypeR
5
Platinum
Rhodium
TypeS
6
Platinum
Rhodium
NOTE