Table 91. api commands, parameters, and arguments – Campbell Scientific CR1000 Measurement and Control System User Manual
Page 374

Section 8. Operation
374
and arguments and the commands wherein they are used. Parameters and
arguments for specific commands are listed in the following sections.
Table 91. API Commands, Parameters, and Arguments
Parameter
Commands in which the
parameter is used
Function of parameter
Argument(s)
uri
•
BrowseSymbols
•
DataQuery
•
ClockSet
•
ClockCheck
•
ListFiles
Specifies the data source.
•
source: dl
(datalogger is data
source): default,
applies to all
commands listed in
column 2.
•
tablename.fieldname
:
applies only to
BrowseSymbols,
and DataQuery
format
•
BrowseSymbols
•
DataQuery
•
ClockSet
•
ClockCheck
•
FileControl
•
ListFiles
Specifies response format.
•
html
,
xml
,
json
:
apply to all
commands listed in
column 2.
•
toa5
and
tob1
apply
only to DataQuery
mode
DataQuery
Specifies range of data with
which to respond.
•
most-recent
•
since-time
•
since-record
•
data-range
•
backfill
p1
DataQuery
•
maximum number of
records (when using
most-recent
argument).
•
beginning date and/or
time (when using
since-time
,or
date-
range
arguments).
•
beginning record
number (when using
since-record
argument).
•
interval in seconds
(when using
backfill
argument).
•
integer number of
records (when using
most-recent
argument)
•
time in defined
format (when using
since-time
,or
date-
range
arguments, see
Time Syntax
(p. 375)
section)
•
integer record
number(when using
since-record
argument).
•
integer number of
seconds (when using
backfill
argument).