10 time skew between measurements – Campbell Scientific CR1000 Measurement and Control System User Manual
Page 294

Section 8. Operation
294
Table 60. Calibrate() Instruction Results
Descriptions of Array Elements
Array
Cal()
Element
Differential (Diff)
Single-Ended (SE)
Offset or Gain
±mV Input
Range
Integration
Typical Value
27
Gain
250
60-Hz
Rejection
-0.067
mV/LSB
28
SE
Offset
25
60-Hz Rejection
±5 LSB
29
Diff
Offset
25
60-Hz Rejection
±5 LSB
30
Gain
25
60-Hz
Rejection
-0.0067
mV/LSB
31
SE
Offset
7.5
60-Hz Rejection
±10 LSB
32 Diff
Offset
7.5 60-Hz Rejection
±10 LSB
33
Gain
7.5
60-Hz
Rejection
-0.002
mV/LSB
34
SE
Offset
2.5
60-Hz Rejection
±20 LSB
35 Diff
Offset
2.5 60-Hz Rejection
±20 LSB
36
Gain
2.5
60-Hz
Rejection
-0.00067
mV/LSB
37
SE
Offset
5000
50-Hz Rejection
±5 LSB
38
Diff
Offset
5000
50-Hz Rejection
±5 LSB
39
Gain 5000
50-Hz
Rejection
-0.67
mV/LSB
40
SE
Offset
2500
50-Hz Rejection
±5 LSB
41
Diff
Offset
2500
50-Hz Rejection
±5 LSB
42
Gain 2500
50-Hz
Rejection
-0.34
mV/LSB
43
SE
Offset
250
50-Hz Rejection
±5 LSB
44
Diff
Offset
250
50-Hz Rejection
±5 LSB
45
Gain
250
50-Hz
Rejection
-0.067
mV/LSB
46
SE
Offset
25
50-Hz Rejection
±5 LSB
47
Diff
Offset
25
50-Hz Rejection
±5 LSB
48
Gain
25
50-Hz
Rejection
-0.0067
mV/LSB
49
SE
Offset
7.5
50-Hz Rejection
±10 LSB
50 Diff
Offset
7.5 50-Hz Rejection
±10 LSB
51
Gain
7.5
50-Hz
Rejection
-0.002
mV/LSB
52
SE
Offset
2.5
50-Hz Rejection
±20 LSB
53 Diff
Offset
2.5 50-Hz Rejection
±20 LSB
54
Gain
2.5
50-Hz
Rejection
-0.00067
mV/LSB
8.1.2.10 Time Skew Between Measurements
Time skew between consecutive voltage measurements is a function of settling
and integration times, A/D conversion, and the number entered into the
Reps
parameter of the VoltDiff() or VoltSE() instruction. A close approximation is:
Time Skew = Settling Time + Integration Time + A‐D Conversion Time
1
+
Reps/NoReps
2