2 measurement rate: 101 to 600 hz – Campbell Scientific CR1000 Measurement and Control System User Manual
Page 233

Section 7. Installation
233
BeginProg
Scan
(1,Sec,0,0)
'<<<
VoltSe
(FastSE(),1,mV2_5,1,False,100,250,1.0,0)
CallTable
FastSETable
NextScan
EndProg
By modifying the
Interval
,
Units
,
and Buffers
arguments, VoltSE() can be
executed at 100 Hz at 100% duty cycle. The following program measures 16
analog input channels at 100 Hz.
Table 34. CRBasic EXAMPLE. Measuring VoltSE() at 100 Hz
PipeLineMode
'<<<
Public
FastSE(16)
DataTable
(FastSETable,1,-1)
Sample
(16,FastSE(),FP2)
EndTable
BeginProg
Scan
(10,mSec,10,0)
'<<<
VoltSe
(FastSE(),16,mV2_5,1,False,100,250,1.0,0)
CallTable
FastSETable
NextScan
EndProg
7.8.12.2 Measurement Rate: 101 to 600 Hz
To measure at rates between 100 and 600 Hz, the SubScan() / NextSubScan
instruction pair is added. Measurements over 100 Hz do not do not have 100%
duty cycle, but are accomplished through measurement bursts. Each burst lasts
for some fraction of the scan interval. During the remainder of the scan interval,
the CR1000 processor catches up on overhead tasks and processes data stored in
the buffers. For example, the CR1000 can be programmed to measure VoltSE()
on 8 channels at 200 Hz with a 95% duty cycle as follows:
Table 35. Measuring VoltSE() at 200 Hz
PipeLineMode
'<<<
Public
BurstSE(8)
DataTable
(BurstSETable,1,-1)
Sample
(8,BurstSE(),FP2)
EndTable
BeginProg
Scan
(1,Sec,10,0)
'<<<
SubScan
(5,mSec,190)
'<<<
VoltSe
(BurstSE(),8,mV2_5,1,False,100,250,1.0,0)
CallTable
BurstSETable
NextSubScan
NextScan
EndProg