beautypg.com

1 zero or tare (option 0), Figure 48: zero (option 0), Table 22. calibration report for air rh sensor – Campbell Scientific CR1000 Measurement and Control System User Manual

Page 154

background image

Section 7. Installation

154 

 

 

"offset" = "y‐ intercept" = "zero" 

"multiplier" = "slope" = "gain" 

7.8.1.5.1 Zero or Tare (Option 0)

Zero option simply adjusts a sensor's output to zero. It does not affect the
multiplier.

Case: A sensor measures the relative humidity (RH) of air. Multiplier is known to
be stable, but sensor offset drifts and requires regular zeroing in a desiccated
chamber. The following procedure zeros the RH sensor to obtain the calibration
report shown. Use the external keyboard / display or software numeric monitor to
change variable values as directed.

Table 22. Calibration Report for Air RH Sensor

Parameter Argument

at

Deployment

Argument at 30-Day

Service

mV output

1000 mV

1050 mV

KnownRH (desiccated
chamber)

0 %

0 %

Multiplier

0.05 % / mV

0.05 % / mV

Offset

-50 %

-52.5 %

RH reading

0 %

0 %

1. Send CRBasic example FieldCal Zeroing Demonstration Program

(p. 155)

to the

CR1000. An excitation channel has been programmed to simulate a sensor
output.

2. To place the simulated RH sensor in a simulated-calibration condition (in the

field it would be placed in a desiccated chamber), place a jumper wire between
channels VX1/EX1 and SE6 (3L). Set variable mV to 1000. Set variable
KnownRH to 0.0.

3. To simulate a calibration, change the value in variable CalMode to 1 to start

calibration. When CalMode increments to 6, zero calibration is complete.
Calibrated Offset will equal -50% at this stage of this example.

Figure 48: Zero (Option 0)

4. To continue this example and simulate a zero-drift condition, change variable

mV to 1050.