Campbell Scientific CR1000 Measurement and Control System User Manual
Page 158

Section 7. Installation
158
Calibration Report for Pressure Transducer
Parameter
Measurement
Before Zero
Measurement
After Zero
Piezometer Output (digits)
8746
0
Piezometer Temperature (°C)
21.4
0
Barometer Pressure (mb)
991
0
1. Send CRBasic example FieldCal() Zero Basis Demo Program
(p. 158)
to the
CR1000.
2. To simulate the pressure transducer in zero conditions:
• Digits_Measured is set to 8746 automatically
• Temp_Measured is set to 21.4 automatically
• BP_Measured to 991
3. To simulate the calibration, change the value in variable CalMode to 1 to start
calibration. When CalMode increments to 6, zero calibrations are complete.
Calibrated offsets will equal 8746 digits, 21.5 C°, and 991 mb.
CRBasic Example 28. FieldCal() Zero Basis Demo Program
'FieldCal zero basis demonstration program
Public
Pressure
Public
VW(1,6)
Public
Equation_Parameters(3)
Alias
Equation_Parameters(1) = Digits_Measured
Alias
Equation_Parameters(2) = Temp_Measured
Alias
Equation_Parameters(3) = BP_Measured
Public
Offset(3)
Alias
Offset(1) = Digits_Offset
Alias
Offset(2) = Temp_Offset
Alias
Offset(3) = BP_Offset
Public
LoadResult, CalMode
Public
AVWRC
Const
GageFactor = 0.01664
Const
Temp_K = -0.00517
BeginProg
'Load the calibration constants stored in the CAL file after a zero is performed
LoadResult = LoadFieldCal(False)
Scan
(1,Sec,1,0)