beautypg.com

19 running average, Crbasic example 61. pt100 in four-wire full-bridge – Campbell Scientific CR1000 Measurement and Control System User Manual

Page 263

background image

Section 7. Installation

263

 

Figure 81: PT100 in four-wire full-bridge

CRBasic Example 61. PT100 in Four‐Wire Full‐Bridge 

'See FIGURE. PT100 in Four-Wire Full-Bridge

(p. 263)

for wiring diagram.

Public

BrFullOut

Public

Rs_Ro

Public

Deg_C

BeginProg

Scan

(1,Sec,0,0)

'BrFull(Dst,Reps,Range,DfChan,Vx1,MPS,Ex,RevEx,RevDf,Settle,Integ,Mult,Offset)

BrFull

(BrFullOut,1,mV25,1,Vx1,1,2500,True,True,0,250,.001,.02344)

'BrTrans = Rf*(X/(1-X))

Rs_Ro = 50 * (BrFullOut/(1 - BrFullOut))

'PRTCalc(Destination,Reps,Source,PRTType,Mult,Offset)

PRTCalc

(Deg_C,1,Rs_Ro,2,1.0,0)

NextScan

EndProg

 

7.8.19 Running Average

The AvgRun() instruction calculates a running average of a measurement or
calculated value. A running average is the average of the last N values where N is
the number of values, as expressed in figure Running-Average Equation

(p. 263),

Figure 82: Running-average equation