beautypg.com

Figure 83: running-average frequency response – Campbell Scientific CR1000 Measurement and Control System User Manual

Page 265

background image

Section 7. Installation

265

 

Note N = Number of points in running average)

To calculate the delay in time, multiply the result from the above equation by the
period at which the running average is executed (usually the scan period):

Delay in time = (scan period) (N ‐ 1) / 2 

For the example above, the delay is:

Delay in time = (1 ms) (4 ‐ 1) / 2 = 1.5 ms 

Example:

Actual test using an accelerometer mounted on a beam whose resonant 
frequency is about 36 Hz. The measurement period was 2 ms. The 
running average duration was 20 ms (frequency of 50 Hz), so the 
normalized resonant frequency is, 

36/50 = 0.72,  SIN(0.72π) / (0.72π) = 0.34. 

The recorded amplitude for this example should be about 1/3 of the 
input‐signal amplitude.  A program was written with two stored 
variables: Accel2 and Accel2RA.  The raw measurement was stored in 
Accel2, while Accel2RA was the result of performing a running average 
on the Accel2 variable. Both values were stored at a rate of 500 Hz. 
Figure Running‐Average Signal Attenuation

 (p. 266

show the two values 

plotted in a single graph to illustrate the attenuation (the running‐
average value has the lower amplitude). 

The resultant delay (delay in time) = (Scan rate)(N‐1)/2 = 2 ms (10‐1)/2 = 
9 ms.  This is about 1/3 of the input‐signal period. 

Figure 83: Running-average frequency response